ATTO3-R High Vacuum Glass Bell Cover Thermal Evaporation Coating Machine

ATTO3-R is a proven, robust and versatile design standard solution. The system can be operated manually and automatically, with a wide range of deposition instrument options, including: RF and DC sputtering sources, low temperature organic evaporators and metal evaporators. There are also many system options available, such as quartz crystal monitoring, high vacuum load lock and heating, cooling, rotating or static sample stage.



Request Quote

FREE:888-999-6186
CALL:541-666-2111

ATTO3-R High Vacuum Glass Bell Cover Thermal Evaporation Coating Machine

Features:

Box 304L stainless steel or glass chamber: Φ300mm × 415mm high (internal dimensions)

The room comes standard with up to five spare ports based on system configuration

Vacuum test runs to 10 -8 Torr

Thermal evaporation (up to 2 independent metal evaporation boats)

Magnetron sputtering source (up to 2 2 ", 3" sources)

Organic deposition sources (up to two)

The above configurations can be installed in combination.

Custom configurations available on request

Technical Parameter: 

Vacuum coating room

Description

Glass bell cover +304 stainless steel vacuum chamber base

Size: Φ300 × H415mm;

Volume

0.03m³

Vacuum system

Description

"Molecular pump + mechanical pump" high vacuum system:

Molecular pump

Pumping speed: 1200L / S, a true quasi-oil-free vacuum system, which uses down pumping to avoid the pumping weakness of the molecular pump and improve the pumping speed. The molecular pump is guaranteed for 5 years and provides direct replacement service for damage.

  * Optional imported molecular pump

Vacuum pump

Specifications: Rotary vane vacuum pump, with oil mist filter, with gas ballast control

Flow: 36 m³ / h (21 cfm), double substrate, vacuum degree

* Optional dry vacuum pump

Operating Voltage

AC 230 V / 50-60 Hz, 10 A or AC 115 V / 50-60 Hz, 20 A (optional)

Pumping speed

Draw from the atmosphere to 9 × 10-4Pa≤30min (expose the atmosphere for a short time, start to evacuate after rushing into dry nitrogen)

Valve

Fore valve / bypass valve: GDQ-40 high vacuum pneumatic baffle valve;

Vacuum measurement

"Two lows and one high" (two resistance gauges measure low vacuum, one ionization gauge measures high vacuum) Digital composite vacuum gauge, measuring range from 1 × 105Pa to 1 × 10-5Pa; it can be displayed and controlled by PC, which can be realized Interlocking operation.

* Optional imported wide-range vacuum transmitter

Vacuum measurement

Partially adopts metal seal and partly adopt fluorine rubber ring seal;

Coating source

Metal evaporation source

Structure

Water-cooled copper electrode + evaporation boat structure, anti-pollution partitions are installed between evaporation sources, and each source is equipped with positioning baffle, intelligent PC control;

Evaporation temperature

Room temperature 0 ~ 1300 ℃

Metal evaporation power supply

Power supply

0 ~ 3000W, the current is regulated by inverter, which is stable and reliable. The power of the power supply is continuously adjustable from 0 to the maximum power.

ControlRS-485 interface, power supply is continuously adjustable from 0 to maximum power.

Number

2 sets, automatic switching controlv

Control system

Description

It adopts windows operating platform and cotrl2000 control system, and adopts IPC + network technology to realize parameter setting of the main components of the whole machine, implement real-time monitoring and intelligent fault diagnosis, and fully automatic monitoring of film thickness. There are two modes of automatic and manual control. Except for taking and putting samples, all other operations are controlled by software on the PC; it provides friendly human-machine operating interfaces such as vacuum system, process settings, and gas filling and deflation systems; on the industrial computer, parameters can be set through the recipe to achieve the process and process of the program. Setting, storage and printing of equipment parameters.

Features

Can effectively solve the coating accuracy, stability and reliability.

Complete program interlocking, complete design and protection against misoperation; system alarms for abnormal conditions such as water shortage, over-current and over-voltage, and corresponding protection measures are implemented.

It adopts rich I / O interface design to fully meet the functional requirements of expansion and external devices.

The display controls the door opening / closing of the coating machine.

Coating process, process, film thickness, convenient and save printing.

Film thickness monitoring and control system

Description

A quartz crystal film thickness controller is used, and a water-cooled film thickness probe is installed near the substrate table, and an industrial computer is connected. The film thickness meter is used to monitor the coating rate and final thickness in real time, and the accuracy can reach ± 1A (0.1nm). The real-time information is fed back to the industrial computer. If the coating thickness reaches the set thickness, the power can be automatically controlled, and the coating can be stopped to achieve the purpose of automatic film thickness control. 

Range

Monitoring thickness range: 1Å ~ 99µ9999Å, resolution 1Å; Monitoring rate range: 0.1Å ~ 9999.9Å.S / s, resolution 0.1Å

Specification

The coating process, process and film thickness settings are all on the PC, which can realize automatic control of the entire process. The coating process can be set on the PC and data can be recorded.

Cooling system

Description

The cooling water circuit is 8 inlets and 8 outlets, and the total water inlet is controlled by a hydraulic relay.

The total inlet and outlet water is connected to a 1P refrigeration cycle water machine, and the temperature control range is 10-25 ℃. Provide stable refrigerated circulating water to targets, metal sources, molecular pumps, and magnetic fluids to ensure stable operation of the equipment.

Other instructions

Product Certification

ISO9001 certification、CE certification、UL certification

Application considerations

Please refer to the danger, warning, caution, etc.